Reagents Setup. Reagents Lock System/Hardware. (HW) Key. • Each ADVIA 560 Lyse reagent. container comes with a reagent. HW key tha... GT Scientific ADVIA 560 & 560 AL Hematology Systems Maximum storage capacity of 100,000 results. Multilingual operating menu enables usability among diverse staff. Bidirectional LIS ... Siemens Healthineers ADVIA 560 Hematology Analyzer System - Medline Manufacturer: SIEMENS HEALTHCARE. Features & Benefits. The ADVIA 560 hematology analyzer is ideal as a stand-alone or backup analy... Medline ADVIA 560 & 560 AL Hematology Systems Maximum storage capacity of 100,000 results. Multilingual operating menu enables usability among diverse staff. Bidirectional LIS ... Siemens Healthineers Advia 560 - Technical Specs | PDF - Scribd [Link]. * ADVIA® 560 and 560 AL Hematology Systems. Technical Specifications. Product Specifications. Technology/Parameters Laser ... Scribd Scalable Hematology Systems - Enviro Meditech Pratama • Runs up to 60 samples per hour. • Processes aspiration volumes as low as 100 μL. • Facilitates efficient manual sampling of. bot... Enviro Meditech Pratama ADVIA 560 and 560 AL Hematology Systems ADVIA 560 and 560 AL Hematology SystemsMaximum output with minimal user intervention. The ADVIA® 560 and 560 AL Hematology Systems... Siemens Healthineers ADVIA® 560 AND 560 AL Hematology Systems Table_title: Specifications Table_content: header: | Manual closed tube volume | 100 µL | row: | Manual closed tube volume: Manual... Cardinal Health ADVIA® 560 AND 560 AL Hematology Systems - Cardinal Health Fast, high-quality CBC testing in any laboratory environment * Runs up to 60 samples per hour. * Processes aspiration volumes as l... Cardinal Health Troobelsooting Wide | PDF | Touchscreen | Valve - Scribd * 1 ADVIA 560 Troubleshooting: System Evaluation 5. * 1.1 ADVIA 560 System Evaluation Checklist . . . . . . . . . . . . . . . . . ... Scribd Hematology Analyzer ADVIA 560 CLIA Non-Waived - McKesson Table_title: Product Specifications Table_content: header: | McKesson # | 1034278 | row: | McKesson #: Manufacturer # | 1034278: 1... McKesson Medical-Surgical 9 sites ADVIA 360, ADVIA 560, and ADVIA 560 AL Systems Jun 9, 2016 —

Report: Siemens ADVIA 560 Hematology System Report Type: User Operation & Technical Summary Subject: Manual Operation, Maintenance, and Troubleshooting Device: Siemens ADVIA 560 (5-Part Differential Hematology Analyzer)

1. Executive Summary The Siemens ADVIA 560 is a fully automated, bench-top hematology analyzer designed for low- to medium-volume laboratories. It provides a Complete Blood Count (CBC), a 5-part white blood cell (WBC) differential, and reticulocyte analysis using laser-based flow cytometry and cyanide-free hemoglobin methodology. While the device is designed for automation, the "manual" aspect of operation covers daily start-up/shutdown procedures, manual sample dilution, manual maintenance modes, and troubleshooting flagged results. 2. Principles of Operation Understanding the manual requires understanding the technology:

WBC/Basophil Counting: Uses a non-lytic method with a specific reagent to count cells based on size and complexity. RBC/PLT Counting: Utilizes a dedicated channel; platelets are separated from RBCs based on volume and refractive index. Hemoglobin (HGB): Measured using a cyanide-free colorimetric method. Differential Count: Uses flow cytometry with a laser diode to create a scattergram (Peroxy vs. Lobularity) to differentiate Neutrophils, Lymphocytes, Monocytes, Eosinophils, and Basophils. advia 560 manual

3. Manual Operation Procedures A. Daily Start-Up (Power-On Procedure)

Power: Turn on the main power switch (usually located on the rear or side panel). Priming: The system will automatically perform a self-check and priming sequence. Warm-up: Allow the laser and fluidics to stabilize (typically 15–30 minutes). QC (Quality Control): Run Level 1 and Level 2 control materials (e.g., 5DIFF and RETIC controls) to ensure the system is within range before processing patient samples.

B. Manual Sample Processing (Open Vial Mode) Reagents Setup

Requirements: Samples must be well-mixed (inverted 8–10 times). The tube must be barcoded or manually assigned an ID. Procedure:

Select "Sample" or "Run" on the main screen. If no barcode is present, manually enter the Patient ID/Accession Number using the external keyboard. Select the desired profile (e.g., CBC + DIFF, or CBC Only). Place the open tube in the sample holder (ensure the probe enters the tube correctly). Close the safety cover/door. Press the "Start" button. The analyzer aspirates the sample (~50 µL).

C. Manual Dilution (Critical Protocol) If a sample exceeds the linear range (e.g., WBC > 200 x 10³/µL), a manual dilution is required. • Each ADVIA 560 Lyse reagent

Procedure:

Prepare the dilution using the specified Siemens Diluent. Standard dilution is usually 1:2 or 1:5 depending on the parameter affected. Mix thoroughly. Process the diluted sample as a normal patient. Note: The operator must manually calculate the final result by multiplying the result by the dilution factor.